Allanalysis
← Back to Articles

Analytical

Gas chromatography/mass spectrometry (GC/MS)

Samples were analyzed using an Agilent 6890N GC system with a 5975 inert XL MS detector and 7683B auto-injector. Separation was done on an HP5MS capillary column (30 m × 0.25 mm, 0.25 μm film). A 1.0 μL sample was injected with a 100:1 split ratio. Helium was used as the carrier gas at 1.2 mL/min. The column temperature was ramped from 150°C to 250°C at 4°C/min, with a 1-minute initial and 5-minute final hold. Electron ionization (70 eV) was used, with injector and transfer line temperatures at 250°C and 260°C, respectively. The scan range was 30–550 m/z. Compounds were identified by comparing retention times with known standards and verifying MS spectra against the NIST library. Samples were analyzed using an Agilent 6890N GC system with a 5975 inert XL MS detector and 7683B auto-injector. Separation was done on an HP5MS capillary column (30 m × 0.25 mm, 0.25 μm film). A 1.0 μL sample was injected with a 100:1 split ratio. Helium was used as the carrier gas at 1.2 mL/min. The column temperature was ramped from 150°C to 250°C at 4°C/min, with a 1-minute initial and 5-minute final hold. Electron ionization (70 eV) was used, with injector and transfer line temperatures at 250°C and 260°C, respectively. The scan range was 30–550 m/z. Compounds were identified by comparing retention times with known standards and verifying MS spectra against the NIST library. Samples were analyzed using an Agilent 6890N GC system with a 5975 inert XL MS detector and 7683B auto-injector. Separation was done on an HP5MS capillary column (30 m × 0.25 mm, 0.25 μm film). A 1.0 μL sample was injected with a 100:1 split ratio. Helium was used as the carrier gas at 1.2 mL/min. The column temperature was ramped from 150°C to 250°C at 4°C/min, with a 1-minute initial and 5-minute final hold. Electron ionization (70 eV) was used, with injector and transfer line temperatures at 250°C and 260°C, respectively. The scan range was 30–550 m/z. Compounds were identified by comparing retention times with known standards and verifying MS spectra against the NIST library.

By Allschoolabs · August 2, 2025 · 19 views

 Gas chromatography/mass spectrometry (GC/MS)
Samples (digested) were analyzed by gas chromatography/mass spectrometry (GC/MS), using Agilent-Technologies (Little Falls, CA, USA) 6890N Network GC system, equipped with an Agilent-Technologies 5975 inert XL Mass selective detector and Agilent- Technologies 7683B series auto injector. The separation was performed on Agilent Technologies capillary column HP5MS (30 m × 0.25 mm; film thickness 0.25 μm). Sample volume of 1.0 μL was injected into the column with split ratio 100:1. The carrier gas used was Helium at a flow rate of 1.2 mL min-1. The column temperature was programmed from 150oC to 250 oC at a linear ramp rate of 4 oC min-1, while the initial and final hold up time was 1 and 5 min, respectively. An electron ionization mode, with ionization energy of 70 eV, was used for GC/MS detection. Injector and MS transfer line temperature were set at 250 oC and 260 oC, respectively. The scanning mass range was selected from 30-550 m/z (mass-to-charge ratio). The unknown samples were identified on the basis of matching of their relative retention times with those of standards (Sigma Chemical Co., St Louis, MO, USA). Samples were further identified and authenticated using their MS spectra compared to those from the NIST mass spectral library of the GC/MS system.
Gas chromatography/mass spectrometry (GC/MS)Analysis.

Comments & Questions (0)

Related Articles

Quality Control Through Analytical Testing
Quality Control Through Analytical Testing

In today's competitive industries, delivering consistent, high-qual...

Metal Composition Testing Using XRF
Metal Composition Testing Using XRF

Knowing the exact composition of a metal is essential in industries...

Failure Analysis in Manufacturing Industries
Failure Analysis in Manufacturing Industries

No manufacturer wants a product to fail after it reaches a customer.

The Importance of Material Characterization
The Importance of Material Characterization

Material characterization is a critical process in scientific resea...

Free Research

Research Journals

Access 25,000,000 scientific papers, medical journals and analytical research.

Open Journal

Need a lab test?

Browse verified tests and book directly from Allanalysis.

Browse Tests
Need help? Choose a contact option
Chat now