Scanning Electron Microscope – Energy Dispersive X-ray Spectroscopy (SEM-EDS)

Sample Preparation Samples were mounted on aluminum stubs using sticky carbon tape and coated with gold for insulation and conductivity. Each sample was labeled and oven-dried at 60°C. To purge the chamber, nitrogen was introduced at 50 psi before vacuuming. Instrument Setup The SEM-EDS analysis was performed using a Phenom ProX system (phenomWorld, Eindhoven, Netherlands). The sample stub was inserted into the chamber and the system evacuated to a vacuum of 5 x 10⁻⁵ Pa. A 15 kV accelerating voltage was applied, initiating the electron beam for imaging and elemental analysis. Imaging and Elemental Analysis The electron beam excited atoms on the sample surface, generating characteristic X-rays. These were detected by the EDS system to determine elemental composition. Magnifications ranged from 10x to 5000x, and EDS provided elemental concentrations in molar percentages. Sample Loading and Positioning Samples were securely fastened to prevent attraction to the detector magnets. The stub was inserted into the holder, adjusted using the height ring (2 mm below holder surface) for optimal resolution. The chamber door was carefully operated and locked automatically when the sample was properly inserted. Optical Imaging and Focus Upon door closure, the sample moved to the optical imaging position. Focus and magnification were adjusted via rotary knobs and on-screen sliders. Both coarse and fine focus adjustments were possible. Image Storage Captured images were saved both to a connected USB 2.0 flash drive and locally on the Windows-based system. USB storage was selected via the system settings menu.

By Allschoolabs · August 2, 2025 · 75 views

Add Article Run Test Sell Product
Scanning Electron Microscope – Energy Dispersive X-ray Spectroscopy (SEM-EDS)
The Scanning Electron Microscope energy dispersive X-ray spectroscopy (SEM-EDS) Phenom Prox model, manufactured by phenomWorld Eindhoven, Netherlands was used to carry out the morphology analysis. Sample is placed on the Aluminium holder stub using sticky carbon tape. The sample was insulated using gold and then grounded electrically. The samples each are then labeled on their stub, then dried in the oven at 60oC. Nitrogen line was opened at 50 psi and the vent button is pressed to fill the area with nitrogen for proper purging of the chamber. The sample holder stub was then placed in the sample chamber holes and the door was shut and the rotary pump picked and a vacuum of 5 x 10-5 Pa was created. The filament light was switched on and the monitor too automatically switched on. At this stage, the accelerator voltage was 15kV and the filament burned out. The atoms on the surface are excited by the electron beam, emitting specific wavelengths of X-rays that are characteristic of the atomic structure of the elements. An energy dispersive detector (a solid-state device that discriminates among X-ray energies) can analyze these X-ray emissions. Appropriate elements are assigned, yielding the composition of the atoms on the specimen surface. The lowest scan mode of 10x is picked and the TV scan clicked. The magnification is then taking to 1000x at a slow scan, 2000, 3000 to 5,000. The Energy dispersion spectrum scan on the intensity of each of the element present and gives the molar concentration in %, then Image was saved. This procedure is called The Scanning Electron Microscope Energy dispersive X-ray spectroscopy (SEM-EDS) and is useful for analyzing the composition of the surface of a specimen

Sample Preparation

Metallic Samples have a certain affinity for the electromagnets within the SEM. This affinity may be strong enough to pull the sample from the stub and up into the detector region. This will cause degradation of imaging capability and will require a service technician to remove this material from the detector. To avoid this scenario, the sample was firmly fastened to the SEM stub

Loading samples

It was ensured that the sample was properly mounted and immobilized on the stub. The height adjustment ring of the sample holder was turned counter-clockwise until the mounting surface is in the highest position and the stub pin was inserted into the hole on the mounting surface, using the tweezers. Also, it was ensured that the stub was inserted in such way that the flat of the stub is seated on the mounting surface. The sample was lowered by turning the height adjustment ring clockwise the sample was positioned correctly as it was at 2 mm below the top surface of the holder, since each one of the vertical marks on the adjustment ring corresponds to 0.5 mm, thus, rotating the adjustment ring by 4 marks lowered the sample by 2 mm. The best resolution was obtained when the sample is positioned 2 mm below the holder surface. The holder surface allows the user to optimize the trade-off between maximum field-of-view (minimum magnification) and image resolution (maximum useful magnification).

The door was opened by pushing the handle upward. The handle was held and the door was raised to its fullest extent and the sample was inserted into the sample holder slot. The sample holder was inserted correctly when the SAMPLE LED lights up green and the message ‘Please load sample’ disappears from the Image screen. The door was closed by sliding it down firmly with little initial force. The door locked automatically when a‘ lights up orange came up, then the sample is loaded and ready for imaging. The Phenom was re-activated and the sample automatically moved to the optical imaging position without sample holder inserted OK button was clicked to confirm The sample holder was then activated and its name was displayed when the holder is inserted.

Optical imaging

After the Phenom door was closed, the sample was transferred automatically to the optical imaging position. The optical camera was activated and the image was displayed in the main viewing window of the Image screen. The part of the sample that was magnified in the main viewing window was displayed in the optical overview window.

Adjusting focus

Touch to activate focus adjustment.

A focus slider appeared, showing the current focus setting. Rotate the rotary knob was rotated to adjust the focus of the optical image. Adjustment was made visible by the slider. the rotary knob was then pressed to select fine focus. An ‘F’ appears on the button and focus adjustment then took in small steps. The rotary knob was pressedagain to return to normal (coarse) focus.

Adjusting magnification (ProX / Pro)

A magnification slider showing the current magnification setting was pressed and the rotary knob was pressed to adjust the magnification of the optical image. The adjustment was made visible by the slider.

Storing images

Images was stored on a USB 2.0 flash drive (USB Flash drive) as well as on Windows.

Selecting storage location

When a USB Flash drive is inserted into one of the Phenom USB ports from SETTINGS in the screen selection bar, the Settings screen appears and the USB was cliked then the images was stored on the USB 2.0 flash drive.

Comments & Questions (0)

Related Articles

Why Blood Tests Are Important for Preventive Healthcare and Early Detection
Why Blood Tests Are Important for Preventive Healthcare and Early Detection

Blood tests can support early detection by identifying changes in bloo…

SEM-EDS Analysis: How Researchers Identify Elements on Material Surfaces
SEM-EDS Analysis: How Researchers Identify Elements on Material Surfaces

SEM-EDS combines scanning electron microscopy with energy-dispersive X…

How to Read Common Blood Test Results Without Getting Confused by the Numbers
How to Read Common Blood Test Results Without Getting Confused by the Numbers

To read common blood test results, check the test name, measured value…

Malaria Testing in Nigeria: When to Test and Which Laboratory Tests Are Used
Malaria Testing in Nigeria: When to Test and Which Laboratory Tests Are Used

Malaria testing in Nigeria is mainly performed using a rapid diagnosti…

Free Research

Research Journals

Access 25,000,000 scientific papers, medical journals and analytical research.

Open Journal

Need a lab test?

Browse verified tests and book directly from Allanalysis.

Browse Tests
Need help?Chat with us or
or contact us
Chat now