QUANTACHROME NOVA4200e Sample Prep & Analysis – Summary Sample Prep: Weigh sample into cell. Insert cell into heating mantle and secure. Connect to prep station, tighten ring. Set degas temp to 250°C, degas for 3 hrs. Cool, remove, and reweigh sample. Sample Analysis: Fill dewar with liquid nitrogen. Insert degassed sample into analysis port. Complete Sample, Points, and Equilibrium menus. Use defaults for BET if needed. Click Start to begin analysis. QUANTACHROME NOVA4200e Sample Prep & Analysis – Summary Sample Prep: Weigh sample into cell. Insert cell into heating mantle and secure. Connect to prep station, tighten ring. Set degas temp to 250°C, degas for 3 hrs. Cool, remove, and reweigh sample. Sample Analysis: Fill dewar with liquid nitrogen. Insert degassed sample into analysis port. Complete Sample, Points, and Equilibrium menus. Use defaults for BET if needed. Click Start to begin analysis.

By Allschoolabs · August 2, 2025 · 42 views

Add Article Run Test Sell Product
 Brunauer–Emmett–Teller (BET)
Sample Preparation.

QUANTACHROME NOVA4200e Made in USA.

Properly weigh the sample into the sample cell. Insert the filled sample cell bulb into the heating mantle. Place the clamp around the mantle so that the sample cell is held firm. Insert the sample cell stem into the sample preparation station. Tighten the knurled ring by turning it clockwise to secure the sample cell in the preparation station. After the sample cell(s) secured in the preparation station, initiate out gassing by entering the control panel menu on the instrument, set the outgassing temperature to 2500c and instruct the system to start degassing for 3hr and switch on the heater. When the sample has out gassed for 3hour’s turn off the heating mantle, Allow the heating mantle and the sample cell cool. Once the heating mantle has cooled, unload the out gas station and remove the samples. Reweigh the sample cell to determine the post out gas sample weight. Sample Analysis

Fill dewar to the internal upper mark with liquid nitrogen, place the sample cell containing an out gassed and weighed sample into the analysis station to be used for analysis! Complete all the fields/selections on the start analysis ‘’Sample’’ menu. Also complete point selection and tagging on the start Analysis ‘’Points’’ menu! Complete fields on the start Analysis ‘’Equilibrium’’ menu! Field defaults can be used for BET measurements! When all fields are completed on the start Analysis Menu, Click ‘’Start’’ to begin the analysis.

Comments & Questions (0)

Related Articles

HbA1c Test Explained
HbA1c Test Explained

The HbA1c test is an important laboratory test that provides an est...

Blood Sugar Test Explained
Blood Sugar Test Explained

Blood sugar testing is an important part of healthcare and can help...

Hematocrit Explained: What It Measures and What Your Results Mean
Hematocrit Explained: What It Measures and What Your Results Mean

Hematocrit is an important blood measurement that shows the proport...

Free Research

Research Journals

Access 25,000,000 scientific papers, medical journals and analytical research.

Open Journal

Need a lab test?

Browse verified tests and book directly from Allanalysis.

Browse Tests
Need help?Chat with us or request a test
or contact us
Chat now