QUANTACHROME NOVA4200e Sample Prep & Analysis – Summary Sample Prep: Weigh sample into cell. Insert cell into heating mantle and secure. Connect to prep station, tighten ring. Set degas temp to 250°C, degas for 3 hrs. Cool, remove, and reweigh sample. Sample Analysis: Fill dewar with liquid nitrogen. Insert degassed sample into analysis port. Complete Sample, Points, and Equilibrium menus. Use defaults for BET if needed. Click Start to begin analysis. QUANTACHROME NOVA4200e Sample Prep & Analysis – Summary Sample Prep: Weigh sample into cell. Insert cell into heating mantle and secure. Connect to prep station, tighten ring. Set degas temp to 250°C, degas for 3 hrs. Cool, remove, and reweigh sample. Sample Analysis: Fill dewar with liquid nitrogen. Insert degassed sample into analysis port. Complete Sample, Points, and Equilibrium menus. Use defaults for BET if needed. Click Start to begin analysis.

By Allschoolabs · August 2, 2025 · 38 views

Add Article Run Test Sell Product
 Brunauer–Emmett–Teller (BET)
Sample Preparation.

QUANTACHROME NOVA4200e Made in USA.

Properly weigh the sample into the sample cell. Insert the filled sample cell bulb into the heating mantle. Place the clamp around the mantle so that the sample cell is held firm. Insert the sample cell stem into the sample preparation station. Tighten the knurled ring by turning it clockwise to secure the sample cell in the preparation station. After the sample cell(s) secured in the preparation station, initiate out gassing by entering the control panel menu on the instrument, set the outgassing temperature to 2500c and instruct the system to start degassing for 3hr and switch on the heater. When the sample has out gassed for 3hour’s turn off the heating mantle, Allow the heating mantle and the sample cell cool. Once the heating mantle has cooled, unload the out gas station and remove the samples. Reweigh the sample cell to determine the post out gas sample weight. Sample Analysis

Fill dewar to the internal upper mark with liquid nitrogen, place the sample cell containing an out gassed and weighed sample into the analysis station to be used for analysis! Complete all the fields/selections on the start analysis ‘’Sample’’ menu. Also complete point selection and tagging on the start Analysis ‘’Points’’ menu! Complete fields on the start Analysis ‘’Equilibrium’’ menu! Field defaults can be used for BET measurements! When all fields are completed on the start Analysis Menu, Click ‘’Start’’ to begin the analysis.

Comments & Questions (0)

Related Articles

Understanding Laboratory Reference Ranges
Understanding Laboratory Reference Ranges

A laboratory reference range is the expected range of values for a ...

What Happens During a Blood Sample Collection?
What Happens During a Blood Sample Collection?

Blood contains important information about how your body is functio...

How to Read Basic Laboratory Reports
How to Read Basic Laboratory Reports

Although only a qualified healthcare provider can interpret your re...

Laboratory Testing and Men's Health
Laboratory Testing and Men's Health

Laboratory tests provide valuable information that cannot be obtain...

Free Research

Research Journals

Access 25,000,000 scientific papers, medical journals and analytical research.

Open Journal

Need a lab test?

Browse verified tests and book directly from Allanalysis.

Browse Tests
Need help? Choose a contact option
Chat now